Thermocouple Calibrator
Thermocouple Calibrator

1.Output of DC mV and simulation of thermocouple output 2.Voltage, DC mV and thermocouple measurement. 3.Manual stepping Basic Introduction  Function

1. Output of DC mV and simulation of thermocouple output.

2. Voltage, DC mV and thermocouple measurement.

3. Automatic and manual cold end temperature compensation for TC.

4.Manual stepping, automatic stepping, 0~100% phase step and slope output

Summary of Source and Measure Functions 

Function

Measure

Source

DC V

0~30V

Not available

DC mV

0~120mV

0mV~100mV

Resistance

0~3200Ω

0~3200Ω

RTD

Pt100, Pt1000, Cu50, Cu100

Pt100, Pt1000, Cu50, Cu100

TC

E,J,K,T,B,R,S,N

E,J,K,T,B,R,S,N

Others

Stepping output, ramp, phase step output, user-defined range

Terminal Description Using Thermocouple (TC)  

The calibrator supports 8 types of standard thermocouples, including J, K, T, E, R, S, B, N.

Lower table summarizes the ranges and characteristics of the supported thermocouples:

Graduation

Positive lead material

Positive wire (H) color

Negative lead material

Specified span°C

ANSI*

IEC**

E

Chrome-nickel

Mauve

Purple

Constantan

-200~950

N

Nickel-chrome-silicon

Orange

Pink

Nickel-silicon-magnesium

-200~1300

J

Iron

White

Black

Constantan

-200~1200

K

Chrome-nickel

Yellow

Green

Alumel

-200~1370

T

Copper

Blue

Brown

Constantan

-200~400

B

Platinum (30% rhodium)

Grey

Platinum (6% rhodium)

600~1800

R

Platinum (13% rhodium)

Black

Orange

Platinum

-20~1750

S

Platinum (10% rhodium)

Black

Orange

Platinum

-20~1750

* Negative conductor (L) regulated as American National Standard (ANSI) is always red.

*Negative conductor (L) regulated as International Electro-technical Commission (L) is always white.

Automatic compensation  

First entered the thermocouple measurement mode, the default of the cold end temperature compensation is automatic compensation reading as shown in Figure 4.1-1.The  sign represents that the current cold end temperature compensation is automatic compensation. If you need further view current automatic cold end temperature compensation value,you have to operation  key. Reading as shown in Figure 4.1.1-1,after press  key, the  sign is replaced for current automatic cold end temperature compensation value as . The value  continue to display for 2s and then automatically return to the  sign. 

Factory Reset  

Factory reset consists of the following items:

1)  The working status recovers to the voltage measurement.

2)   Automatic shutdown time is reset to be 30 min, which is effective.

3)    LCD backlight brightness is set to be 60%.

4)   Output span is recovered to be factory default.

Start the calibrator and press  until the buzzer works, and the recovery of factory default will enter working mode when the recovery is completed.