Welcome to ECBAY
Automated Elisa analyzer
Automated Elisa analyzer

Fully Automated Elisa analyzer, test ELISA, TP, HIV, HCV,T3,T4 etc.

STANDARD CONFIGURATION

Specifications

ADC ELISA200

Pipetting Spec

Continuous load sample, parallel serial pipetting

Sample and reagent pipetting Probe

2pc

Tip type

Disposable plastic tips, non conductive, low cost

Liquid level, clot, empty tube detection function

Sample-tip range

Tip type 300µL,800µL optional

Tips rack capacity

384tips(4positions, each can hold 96tips)

Replicates

Up to 120 samples, standards and controlsPrecision, tip:<3CV,

Accuracy

≤2.0%( 100µl)

Precision

≤0.75%( 100µl)

Reagent fluid capacity

Standard:24reagent tanks position

Intelligent transfer arm

1pc, grab the plate tightly, avoid dropping

Incubator Spec

Incubation plates

4pc, 1pc RT position, each plate up to 12 assays

Temperature range

Ambient +5 oC--60 oC   uniformity ±0.5oC

Shaking, temperature control

3 Independent shaking , parameters can be set

Washer Spec

Manifold configuration

8-pin washer

Dispense-volume range

50-1000µl, residual volume<2µL

Cross washing function

Yes, liquid inject height, soak time etc multi-parameter programmable

Wash buffer capacity

3x 2L Low-buffer alarm function

Reader Spec

Dynamic range

0-4.0 OD  

Spectral range

340~750 nm

Filter shots

4 standard configurations(405/450/492/630) 2 vacancy optional

Reading channels

Linear scan, simultaneously read 8 wells.

Reading modes

Single, dual   CV<3%

HIS/LIS, CE certificate, ISO9001,ISO13485, Completely intellectual property rights

   

Automated Elisa analyzer