Patient Simulator 3.5" TFT display led backlight 320x240 pixels,touch screen ECG simulation/ 12 leads outputs rechargable

Multi-parameter Patient Simulator

MODEL: MS400

 

 

1.3.5-inch color TFT touch-screen, 320*240 pixels, friendly operation interface.

2.Display the output waveforms synchronously, which is convenient for user to

 compare with the test equipment.

3.It integrates the functions of SD socket interface, user can output the

  custom waveform by this device.

4.Built-in templates list to provide the waveform that is expressed by each

 preset sequence number,which is convenient for searching.

5.Two operation interfaces:Chinese and English.

6.Operate with keystroke and touch-screen expediently and rapidly.

7.Built-in 3.7V/4000mAh(Rechargeable alkaline battery).

8.Optical external power: AC 220V, 50Hz or AC 120V, 60Hz

ECG/Arrhythmia

ECG

Conventional: All leads have the same voltage with the II-lead as the reference

Amplitude: 0.5,1.0,1.5 and 2.0mV

The accuracy of Amplitude: ±2%

ECG Rate:

 15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340 and 350 BPM

ECG artifacts: 50 or 60 Hz, muscle, baseline drift and breathing

The elevation of the ST segment: -0.8 mV to +0.8 mV, the step is 0.1 mV

The default: 80 BPM,1mV

Arrhythmia

Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia,

 Missed Beat, Atrial

Tachycardia, Nodal Rhythm, Supraventricular Tachycardia

Ventricular: Vent Fib (coarse), Vent Fib (fine), Premature Ventricular

  Contraction (PVC) 6/min,

PVCs 12/min, PVCs 24/min, Freqent Multifocal, Bigeminy, Trigeminy, Paired

  PVCs, Run 5

PVCs, Run 11 PVCs, Vent Tachycardia, Asystole

Premature Beat: Atrial Premature Atrial Contraction (PAC), Nodal PNC, Left

 Ventricular Focus

PVC1, Early Left Ventricular Focus PVC1, Right Ventricular Focus PVC1, Early

  Right

Ventricular Focus PVC1, R on T Left Ventricular Focus, R on T Right

 Ventricular Focus, Multifocal PVC

Conduction Defects: First Degree Heart Block, Second Degree Heart Block,

  Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch

  Block 

ECG Performance Testing

ECG amplitude:0.5,1.0,1.5 and 2.0mV

The accuracy of Amplitude: ±2%

Pulse: 30 and 60 BPM, the width of pulse wave is 60 ms

Square: 0.125 and 2.0 Hz

Triangle: 2.0 and 2.5 Hz

Sine: 0.5, 5, 10, 40, 50, 60 and 100 Hz

ECG Rate:15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240,

 260, 280, 300,   320, 340 and 350 BPM

The accuracy of ECG Rate: ±1%

Respiration

Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIIN

Respiration Impedance: 0.2, 0.5, 1.0 and 3.0 Ω

Baseline Impedance:500,1000,1500 and 2000

Apnea: 12 s, 22 s, 32 s

ECG simulator with adjustable heart rate/12-lead output

33 kinds of Arrhythmia waveforms

Pacemaker

Two IBP channels

Variety of respiratory impedance,apnea pauses (0 BrPM) of 12, 22 and 32

  seconds.

Standard ECG calibration

Export square,triangle and sine with adjustable frequency

IBP

The default: 00/00mmHg

Dynamic IBP

Channel 1: Artery  120/80mmHg

               Radial 120/80mmHg

               Left Ventricle 120/00mmHg

               Right Ventricle 25/00mmHg

Channel 2: Artery  120/80mmHg

               Radial 120/80mmHg

               Left Ventricle 120/00mmHg

               Right Atrium 15/10mmHg

               Right Ventricle 25/00mmHg

               Pulmonary Artery 25/10mmHg

               Left Atrium 14/4mmHg

Static-Pressure Levels

Channel 1: 0, 80, 160, 240, 320 and 400mmHg

Channel 2: 0, 50, 100, 150, 200 and 250mmHg

Packing

About Us

CONTEC Medical Systems focusing on research, manufacture and distribution of medical equipments, was founded in 1992 as a professional medical equipments producer. Our Products include followings :

ECG / ECG holter           EEG / EEG Holter      B-ultrasound Scanner

Patient Monitor        Fetal doppler             Fetal Monitor 

Spo2 Monitor    Stethoscope              Medical Image

MS400 Multi-parameter Patient Simulator - ECG, Arrhythmia, IBP, respiration, temprature

Ads by Google


About Us | Contact Us | Help | Terms & Conditions
Hot Products: A | B | C | D | E | F | G | H | I | J | K | L | M | N | O | P | Q | R | S | T | U | V | W | X | Y | Z | 0-9
Copyright Notice @ 2008-2022 ECBAY Limited and/or its subsidiaries and licensors. All rights reserved.