1.Single channel ECG machine with measurement 2.Manual/Auto/Arrhythmia Analysis Modes 3.ECG waveform, 16 cases save/replay
Features:
Lead: Standard 12 leadsAcquisition: Simultaneous 12 leadsResolution: 12bit/1000HzOperation mode: Manual/Auto/Analysis ModesFilter: AC.EMG FilterDrift filter: Anti-Drift SystemCMRR: ≥100dB, with AC FilterInput circuit: Floating, Protection against defibrillator effectInput circuit current: ≤0.1µADrift filter: Anti-Drift SystemCMRR: ≥100dB, with AC FilterInput circuit: Floating, Protection against defibrillator effectInput circuit current: ≤0.1µAInput Impedance: >50MΩPatient current leakage: <10µACalibrating voltage: 1mV±3%Voltage tolerance: ±500mVTime constant: >3.2sFrequent response: 0.05Hz~160HzNoise level: ≤15µVp-pThreshold: ≤20µVPaper speed: 25, 50mm/s (±3%)Sensitivity: Auto, 2.5, 5, 10, 20, 40mm/mVRecorder: High resolution thermal printerSafety standard: GB9706.1-2007, IEC Class I, Type CF
Specifications:
Single channel ECG machine with measurement |
Manual/Auto/Arrhythmia Analysis Modes |
Built in rechargable NI-MH battery 12V (1500mAh) |
Adapt to 110-230V, 50/60HZ AC power supply |
Built in USB/RS232 interface (ECGNET software optional) |
ECG waveform, 16 cases save/replay |
Package: 40cm×34cm×22cm GW: 4.6KG NW: 1.7KG |
320×240 graphic 3.8inch LCD, 16cases save/replay |
ECG-102 Digital Electrocardiograph